Сегодня 17 ноября 2024
Медикус в соцсетях
 
Задать вопрос

ЗАДАТЬ ВОПРОС РЕДАКТОРУ РАЗДЕЛА (ответ в течение нескольких дней)

Представьтесь:
E-mail:
Не публикуется
служит для обратной связи
Антиспам - не удалять!
Ваш вопрос:
Получать ответы и новости раздела
19 декабря 2001 00:00   |   Giuseppe D'Ancona, MD, Hratch Karamanoukian, MD, Akira Kawaguchi,

Coronary Artery Exposure in Off-Pump CABG: A Word of Caution

 
 
ABSTRACT
Coronary exposure and stabilization have focal importance in 
 
off-pump coronary surgery. Off-pump complete myocardial
revascularization can be performed safely in the majority of the 
 
patients whenever strict surgical protocols are followed. Although
new devices may be used to facilitate the performance of this 
 
demanding operation, technical pitfalls should be recognized to 
 
ensure the success of the procedure. We herein report our timely
experience with the Xpose device (Guidant Corp., Cupertino, CA).
INTRODUCTION
The increasing popularity of off-pump coronary artery bypass
grafting (OPCAB) has stimulated interest and concerns about
different techniques of coronary exposure and stabilization. In our 
 
experience, complete myocardial revascularization may be performed
safely without cardiopulmonary bypass (CPB) and via median
sternotomy if strict surgical protocols for coronary exposure and 
 
stabilization are followed in order to maintain adequate
hemodynamics during the entire procedure. Since March 1998, the 
 
introduction of the «single suture» technique [Bergsland 1999] has 
 
drastically changed the applicability of beating heart coronary
surgery. A single heavy suture is placed in the oblique sinus of the 
 
pericardium after having elevated the heart from the pericardial
cradle. The suture is then passed through a double-armed vaginal
tape and is snared down to the posterior pericardium [Bergsland
1999]. Different degrees of traction on the suture and different
positioning of the vaginal tape allow adequate exposure of the 
 
different coronary branches including the topographically more 
 
difficult circumflex coronary artery. Our experience has shown that 
 
normal hemodynamic values can be maintained during lateral
revascularization in the majority of the patients with the «single
suture» technique [D'Ancona 2000].
Recently a new device (Xpose, Guidant Corp., Cupertino, CA.) has 
 
been developed to enrich the surgical armamentarium for complete
myocardial revascularization off-CPB. The Xpose device consists of 
 
a suction cup system that is placed at the apex of the left 
 
ventricle. The cup is connected to an articulated arm that is fixed
to the sternal retractor. Once suction is applied, the tip of the 
 
heart can be gently elevated and different positions of the arm 
 
allow for exposure of different coronary artery branches. The 
 
clinical applicability of this device has been recently demonstrated
in an elegant study [Dullum 2000]. Although the Xpose is a unique
and user friendly system, some associated technical pitfalls should
be recognized. We herein summarize a timely case report.
CASE REPORT
TA is a 67−year-old male, with three-vessel coronary artery disease,
class I angina pectoris, and preserved ejection fraction. OPCAB was 
 
scheduled electively. After induction of general anesthesia and 
 
endo-tracheal intubation, a median sternotomy was performed. The 
 
left internal mammary artery (LIMA) was harvested together with 
 
saphenous vein conduits. After systemic heparinization, the 
 
pericardium was opened using an inverted T incision. The heart was 
 
normal in size with preserved global contractility. The coronary
targets were adequate for OPCAB [Bergsland 2000]. The Xpose
cup-suction was placed on the apex of the left ventricle to expose
the left anterior descending coronary artery (LAD). The LAD was a 
 
2−mm vessel with 80% proximal stenosis. The vessel extended to the 
 
apex of the ventricle. After mechanical stabilization, placement of 
 
a 4−0 polypropilene pledgetted suture, the coronary was opened and a 
 
2−mm intracoronary shunt was placed to maintain adequate distal
perfusion. The proximal snare was released and 7−0 continuous
polypropilene running suture was used to anastomose the LIMA to the 
 
LAD.
During the procedure, multiple premature ventricular contractions
with ST segment elevation in the anterior leads were noted. At the 
 
end of the anastomosis, intraoperative graft patency verification
was performed using transit time flow measurement (TTFM) (Medi-Stim,
Oslo, Norway). A 2−cm segment of the LIMA was skeletonized and a 
 
3−mm flow probe was placed around the LIMA, with the Xpose in 
 
place. The TTFM curve showed an adequate diastolic pattern with a 
 
relatively low absolute flow value [Figure 1 :1791:]. Initial TTFM 
 
analysis suggested patent LIMA to LAD anastomosis.
While measuring flow, ST segment elevation was accompanied with 
 
sudden hemodynamic impairment. At closer inspection, the Xpose
suction cup seemed to compress the distal segment of the native LAD.
We immediately released the tip of the heart from the suction cup 
 
and reassessed flow within the LIMA graft. Transit time flow 
 
measurements were repeated and a significant increase in the 
 
absolute flow value was noted [Figure 2 :1792:]. The EKG tracing
reverted to normal and the systemic hypotension resolved. After
that, we used the «single suture technique» to achieve adequate
exposure of the remaining coronary targets. The operation was 
 
successful, without perioperative morbidity or mortality.
DISCUSSION
OPCAB is performed successfully using different techniques of 
 
coronary exposure and stabilization. In our experience, simple,
standardized, and reproducible protocols have been used to ensure
satisfactory technical results and postoperative outcomes. Minimal
changes in the surgical routine can drastically compromise the 
 
success of the operation and, for this reason, should always be 
 
adopted cautiously. The «single suture» technique enables coronary
exposure while different degrees of traction are placed on the 
 
pericardium. In theory, no compression on the heart and the 
 
epicardial vessels is utilized with this technique.
Although the Xpose can achieve adequate coronary exposure, direct
snaring of the apical vessels with the suction cup should be 
 
avoided. As documented in this case report, compression on the 
 
coronary artery branches not only reduces native flow causing
regional ischemia, but also limits actual flow through newly
constructed anastomoses. As a consequence, erroneous revisions of 
 
patent grafts can be avoided if the pitfall is recognized. This case 
 
report also documents the wide applicability of intraoperative
flowmetry [Canver 1994, Louagie 1994, Walpoth 1996, Canver 1997,
Jaber 1998a, Jaber 1998b, Louagie 1998, Walpoth 1998, Cerrito 1999,
D'Ancona 1999, Di Giammarco 1999, Walpoth 1999]. This technology may 
 
be useful in detecting anastomotic imperfections [Canver 1994,
Louagie 1994, Walpoth 1996, Canver 1997, Jaber 1998a, Jaber 1998b,
Louagie 1998, Walpoth 1998, Cerrito 1999, D'Ancona 1999, Di 
 
Giammarco 1999, Walpoth 1999] and in identifying modifications in 
 
graft flow during the different phases of the operation. Proper
function of coronary grafts should always be ensured at any time to 
 
maintain adequate hemodynamics during off-pump coronary surgery and 
 
to prevent emergent conversions to CPB.
REVIEW AND COMMENTARY
1. Editorial Board Member SC389 writes:
There are no values such as PI shown with Figure 1. There should be 
 
a picture or diagram of placement of the Xpose device on the heart
so that the location of the device and also the placement of the 
 
heart can be visualized so as to assess incorrect displacement, such
as RV or LV compression.
I am concerned about the conclusion from this case report. I do not 
 
have the familiarity and experience with the TTFM that the authors
obviously have, but just looking at the absolute numbers of flow in 
 
Figure 1 and Figure 2 -- Figure 1 is 33 ml/min and Figure 2 is 7 
 
ml/min. I have significant experience with off pump cases with and 
 
without the Xpose and have seen these changes occur from many other
reasons with spontaneous resolution and feel very strongly that this 
 
conclusion can not be drawn from this case.
I would be interested in knowing if the authors have ever had these
changes occur in any of their prior cases without the Xpose. I am 
 
also interested in exactly where the Xpose was placed to be able to 
 
compress the distal LAD and how much of the LAD was compressed. In 
 
the early Octopus experience, it was shown that a suction device
could be placed directly over a coronary vessel without damage to
the vessel.
Authors' Response by Giuseppe D'Ancona, MD:
PI values were adequate in both measurements. The suction cup was 
 
placed on the apex of the heart and was compressing the terminal
portion of the LAD (last 3 cm). The right pleura was fully opened to 
 
prevent compression on the right ventricle. Measurements were done 
 
after removing the coronary stabilizer to prevent compression on the
left ventricle.
We usually test graft flow many times (an average of 20 times/case)
during the same operation to detect any possible kinking,
compression, or excessive tension on the newly constructed grafts,
especially when the heart is elevated and rotated to expose the 
 
lateral coronary targets. If the heart stays in a fixed position and 
 
there are no sudden changes in blood pressure, we very seldom see 
 
drastic changes in the flow curves and values.
As already proven in the Octopus experience, direct suction does not 
 
cause any sort of endothelial lesion on the coronary vessels. On the 
 
contrary, no one has ever shown any data about how coronary flow may 
 
change during the use of these suction devices. It is logical to 
 
believe that, although the suction cup will not cause any sort of 
 
permanent damage of the vessel, its compressing effect may 
 
temporarily deform the coronary artery and consequently reduce the 
 
blood flow.
2. Editorial Board Member MN393 writes:
The original work on suction stabilization (Grundeman, Borst, et 
 
al.) showed no ill effect of 400 mmHg suction applied directly to 
 
the epicardium over a coronary artery. The present case report
strongly suggests coronary compression. Could this be related to the 
 
design of the Xpose device? I have observed that this draws the 
 
whole apex of the heart into the cup (i.e., curving around the outer
lip). Clearly this increases the amount of traction one can obtain.
However, I can see how a relatively unsupported epicardial coronary
artery (lying in epicardial fat) could be kinked around this edge.
Authors' Response by Giuseppe D'Ancona, MD:
Yes, we believe that a direct compression of the LAD and the 
 
surrounding tissue may be the explanation for our findings. A 
 
suctioning cup with a smaller design may be an option.
3. Editorial Board Member EE455 writes:
To me ischemia to the distal LAD, though likely, has not definitely
proven to be the origin of the problem. In this case report, the 
 
fact that removing the Xpose system improved the situation is of 
 
course convincing in its role as a trigger of hemodynamic
instability, but other factors may have been involved. Further
details should be provided, if available (e.g., TEE findings,
hemodynamic measurments, etc.).
ECG changes are not interpretable and some ESV when the heart is 
 
tilted are frequent; there are no definitive proofs of severe
ischemia. Authors should provide further data about the acute
hemodynamic intolerance that happened after completion of the 
 
anastomosis. In my experience, compression of the LAD with the 
 
Xpose system, if any, is very distal and would be unlikely to 
 
provoke major ischemia. Could any other factor have been an issue,
such as inappropriate twisting of the great vessels, or compromise
of the RV function?
Authors' Response by Giuseppe D'Ancona, MD:
Grafting of the LAD requires only minimal displacement and 
 
manipulation of the heart and, for this reason, hemodynamic
impairment due to compression of the ventricles and twisting of the 
 
great vessels is very seldom. On the contrary, we have often noted
sudden instability for ischemic reasons, for example, when the LAD 
 
is not properly shunted or the vessel is excessively compressed by 
 
the stabilizer foot. In this case report, signs of ischemia were 
 
noted after some minutes from the placement of the Xpose, during
construction of the LAD anastomosis. The coronary was shunted and 
 
there were no evident causes of ischemia. At the end of the 
 
anastomosis, the coronary stabilizer was removed, the mammary was 
 
opened to reperfuse the distal LAD, and the heart was almost back to 
 
its initial position. In spite of this, EKG changes persisted and 
 
were accompanied by low-graft flow and then by sudden hypotension.
Because there were no other explanations, we suspected the Xpose to 
 
be the cause of our findings. Immediately after removal of the 
 
suction cup, the situation reverted to normality. This should be 
 
enough to justify our reasonable doubts.
4. Editorial Board Member PB44 writes:
Was the distal LAD visible epicardially and could this have been 
 
anticipated from the angiogram? The LAD usually needs minimal
displacement to perform an anastomosis, thus a stitch usually is 
 
more than adequate. Should one not reserve devices like this to when
they are best needed?
Authors' Response by Giuseppe D'Ancona, MD:
The LAD was visible epicardially.
We agree that exposure of the LAD and other coronary artery branches
can be easily achieved using different strategies.
AUTHOR/ARTICLE INFORMATION
Address corespondence and reprint requests to: Hratch Karamanoukian,
MD, 100 High Street, Buffalo General Hospital, Buffalo, NY, 14203,
Phone: (716) 859−2248, Fax: (716) 859−4697, EMail: lisbon5@yahoo.com
REFERENCES
1. Bergsland J, Karamanoukian HL, Soltoski PR, Salerno TA. «Single
suture» for circumflex exposure in off-pump coronary artery bypass
grafting. Ann Thorac Surg 68(4):1428−30, 1999.
2. Bergsland J, D'Ancona G, Karamanoukian H, Ricci M, Schmid S,
Salerno T. Technical tips and pitfalls in OPCAB surgery: The Buffalo
experience. Heart Surgery Forum, #2000−7789 3(3):189−93, 2000.
3. Canver CC, Dame N. Ultrasonic assessment of internal thoracic
artery graft flow in the revascularized heart. Ann Thorac Surg 
 
58:135−8, 1994.
4. Canver CC, Cooler SD, Murray EL et al. Clinical importance of 
 
measuring coronary graft flows in the revascularized heart.
Ultrasonic or electromagnetic? J Cardiovasc Surg 38:211−5, 1997.
5. Cerrito PB, Koenig SC, Van Himbergen DJ, Jaber SF, Ewert DL,
Spence PA. Neural network pattern recognition analysis of graft-flow
characteristics improves intra-operative anastomotic error detection
in minimally invasive CABG. Eur J Cardiothorac Surg 16:88−93, 1999.
6. D'Ancona G, Karamanoukian H, Salerno TA, Schmid S, Bergsland J.
Flow measurement in coronary surgery. Heart Surgery Forum,
#1999−03815 2(2):121−4, 1999.
7. D'Ancona G, Karamanoukian H, Lima R, Ricci M, Bergsland J,
Rossman J, Salerno T. The hemodynamic effects of elevation and 
 
stabilization of the heart during off-pump coronary surgery. J 
 
Cardiac Surg 2000 (in press).
8. Di Giammarco G. Myocardial revascularization without
cardiopulmonary bypass. Presented at the simposium: State of the art 
 
in emerging coronary revascularization; EACTS, Glasgow, Scotland,
Sept. 4, 1999.
9. Dullum MKC, Resano FG. Xpose: A new device that provides
reproducible and easy access for multivessel beating heart bypass
grafting. Heart Surgery Forum, #2000−7611 3(2):113−8, 2000.
10. Jaber SF, Koenig SC, BhaskerRao B et al. Role of graft flow 
 
measurement technique in anastomotic quality assessment in minimally
invasive CABG. Ann Thorac Surg 66:1087−92, 1998a.
11. Jaber SF, Koenig SC, BhaskerRao B, Van Himbergen DJ, Spence PA.
Can visual assessment of flow waveform morphology detedt anastomotic
error in off pump coronary artery bypass grafting? Eur J 
 
Cardio-Thorac Surg 14:476−9, 1998b.
12. Louagie YAG, Haxhe JP, Jamarat J, Buch M, JC Schoevaerdts.
Intraoperative assessment of coronary artery bypass grafts using a 
 
pulsed doppler flowmeter. Ann Thorac Surg 58:742−9, 1994.
13. Louagie YAG, Brockmann CE, Jamarat J et al. Pulsed Doppler
intraoperative flow assessment and midterm coronary graft patency.
Ann Thorac Surg 66:1282−8, 1998.
14. Walpoth BH, Bosshard A, Kipfer B et al. Failed coronary artery
bypass anastomosis detected by intraoperative coronary flow 
 
measurement. Eur J Cardio-thorac Surg 10:1064−70, 1996.
15. Walpoth BH, Bosshard A, Genyk I et al. Transit time flow 
 
measurement for detection of early graft failure during myocardial
revascularization. Ann Thorac Surg 66:1097−1100, 1998.
16. Walpoth BH, Muller MF, Genyk I et al. Evaluation of coronary
bypass flow with color-Doppler and magnetic resonance imaging
techniques: comparison with intraoperative flow measurements. Eur J 
 
Cardio-Thorac Surg 15:795−802, 1999.

Поделиться:




Комментарии
Смотри также
19 декабря 2001  |  00:12
Low Ejection Fraction Is Not a Contraindication to Off-Pump
Low Ejection Fraction Is Not a Contraindication to Off-Pump
19 декабря 2001  |  00:12
Totally endoscopic myocardial revascularization (experimental study).
Totally endoscopic myocardial revascularization (experimental study).
19 декабря 2001  |  00:12
Posterior Thoracotomy for Reoperative Coronary Artery Bypass
Posterior Thoracotomy for Reoperative Coronary Artery Bypass
19 декабря 2001  |  00:12
Technical Tips and Pitfalls in OPCAB Surgery: The Buffalo
Technical Tips and Pitfalls in OPCAB Surgery: The Buffalo
18 декабря 2001  |  00:12
Как пробиться на международную арену
Как пробиться на международную арену